Battery Cycle Tester
-
Battery Cell Cycler
-
Battery Cycler System
-
Lithium Ion Battery Analyzer
-
Battery Capacity Discharge Tester
-
18650 Battery Charger Capacity Tester
-
8 Channel Battery Analyzer
-
Battery Cell Tester
-
Regenerative Battery Pack Test System
-
Neware Battery Test System
-
EV Battery Testing Equipment
-
Battery Grading System
-
Battery Test Chamber
-
Mr. SAVIN Jean-FrancoisWe have been using Neware benches for testing our primary cells and batteries for several years. We are quite satisfied about the quality of this type of equipment and we find always good and very reactive feedback from their after-sales service to help us in specific case.
-
Prof. Lv LiWe have been using Neware battery testing system for many years. We are focused on different electrochemical tests of lithium-ion batteries. The Neware machine is easy to be used and the program is easy to set. We have conducted charge-discharge cycling for more than 5000 cycles and the machine showed very good stability, which helped us to obtain
-
Mr. Frank GroenVarious neware testing systems have been equipped our laboratory for years and serves very good. We use them for many kinds of tests. Easy to handle the tests and data is well organized for our reporting work. The most important thing is the measurement and control are quite accurate and reliable.
IP20 Battery Health Tester , Pulse DCIR 10HZ Neware Battery Cycler
Discharge Voltage | -5V | Current Range | 5uA - 1mA 1mA - 5mA 5mA - 10mA |
---|---|---|---|
Brand | Neware | End Condition | Voltage , Current , Relative Time , Capacity |
Dimensions | 480mm – 404mm - 130mm | Protection | Power Down Data Protect |
Loop Range | 65535 Time | IP Protection Level | Protection Level IP20 |
High Light | Pulse DCIR battery health tester,Pulse DCIR Neware Battery Cycler,IP20 neware battery cycler |
Neware 5V10MA Battery Cycler System Pulse DCIR 10HZ Data Recording Tester
Want to use a battery tester for all kinds of purposes , wanna give it a shot with low risk , BTS4000 is very possibly the correct choice for you to get to know Neware and her battery test products better . Till year 2017 , 90% new customers of Neware would choose BTS4000 as a starting point , its really the time - proven and reliable choice .
Due to the rapid data acquisition frequency , BTS4000 could be applied to super capacitor or EDLC - Electronic Double Layer Capacitor - test . Through our advanced software, the capacitance, DCIR Direct Current Internal Resistance could be easily calculated .
Higher Accuracy . More Cost-effective. We believe it can improve your battery test .
Neware Battery Charge and Discharge Test System , ranging from μA single cell applications up to 1MW packs , is a high precision system designed specifically for testing lithium ion battery cells , modules , packs and capacitors . It is suitable for product development , quality control , and helpful to characteristic research , cycle life testing, product screening , and quality assessment .
The lower range is able to output current low as 5uA , which can tell the subtle difference of various materials such electrodes , electrolyte , separator , etc .
Generally , magnetoelectric galvanometer can achieve uA level , electromagnetic galvanometer can achieve mA level .
Sensitive galvanometer is used for students to check weak current or tiny voltage in DC circuit in the lab . It is a kind of high sensitive magnetoelectric instrument , which can measure the tiny current .
resolution | AD 16bit , DA 16bit | |
Input Power | 25W | |
Channel Features | Constant Current Source and Constant Voltage Source with Independent pairs of Closed - loop Structure | |
Channel Control Mode | Independent Control | |
Voltage | Per Channel Voltage Range | 0V - 5V |
Discharge Min Voltage | 0V | |
Accuracy | 0.05 percent of FS | |
Stability | 0.05 percent of FS | |
Current | Per Channel Current Range | Range1 , 5uA – 1mA Range2 , 1mA – 5mA Range3 , 5mA – 10mA |
Power | Per Channel Output Power | 0.25W |
Stop current | Range 1 , 2uA Range 2 , 0.01mA Range 3 , 0.02mA | |
Data Recording | Record Condition | Min Time - 100mS |
Min Voltage - 10mV | ||
Min Current - Range 1 , 2uA Range 2 , 0.01mA Range 3 , 0.02mA | ||
Record Frequency | 10Hz | |
Charge | Charge Mode | CCCharge , CVCharge , CCCVCharge , CPCharge |
End Conditon | Voltage , Current , Relative Time , Capacity , -deltaV | |
Discharge | Discharge Mode | CCD , CPD , CRD |
End Condition | Voltage , Current , Relative Time , Capacity | |
Pulse | charge | CCC , CPC |
Discharge | CCD , CPD | |
Min Pulse Width | 500ms | |
Automated Switch | Automated swith from charge to discharge for each pulse | |
End Condition | Voltage, Test Time | |
Cycle | Loop Measure Range | 65535 times |
Max Steps Per Loop | 254 | |
Nested Loop | Nested loop Function , Max Support 3 Layers | |
Protection | Software Protection | Power - down Data Protection |
Off - line Testing Function | ||
Setting Protection Condition,Setting Parameter - Low Voltage limit , Upper Voltage limit , Low Current limit , Upper Current limit , Delay Time | ||
IP protection level | Protection level IP20 | |
Voltage and Current Testing Sample | 4 - wire connecting | |
Database | MySQL Database | |
A Means of Communication of Upper Machine | TCP / IP Protocol | |
Communication Interface | Ethernet Port | |
Per Unit Main Channels No. | 8 |