Battery Cycle Tester
-
Battery Cell Cycler
-
Battery Cycler System
-
Lithium Ion Battery Analyzer
-
Battery Capacity Discharge Tester
-
18650 Battery Charger Capacity Tester
-
8 Channel Battery Analyzer
-
Battery Cell Tester
-
Regenerative Battery Pack Test System
-
Neware Battery Test System
-
EV Battery Testing Equipment
-
Battery Grading System
-
Battery Test Chamber
-
Mr. SAVIN Jean-FrancoisWe have been using Neware benches for testing our primary cells and batteries for several years. We are quite satisfied about the quality of this type of equipment and we find always good and very reactive feedback from their after-sales service to help us in specific case.
-
Prof. Lv LiWe have been using Neware battery testing system for many years. We are focused on different electrochemical tests of lithium-ion batteries. The Neware machine is easy to be used and the program is easy to set. We have conducted charge-discharge cycling for more than 5000 cycles and the machine showed very good stability, which helped us to obtain
-
Mr. Frank GroenVarious neware testing systems have been equipped our laboratory for years and serves very good. We use them for many kinds of tests. Easy to handle the tests and data is well organized for our reporting work. The most important thing is the measurement and control are quite accurate and reliable.
TWS LIR1654 Button Battery Cycler System For Rechargeable Battery
Frequency | 10Hz | Range1 | 5uA – 1mA |
---|---|---|---|
Range2 | 1mA – 10mA | Range3 | 10mA – 20mA |
Main Channels No. | 8 Each Unit | Max Steps Per Loop | 254 |
High Light | TWS Battery Cycler System,LIR1654 Battery Cycler System,LIR1654 rechargeable battery capacity tester |
Neware TWS LIR1654 Button Battery Cycle Tester Rechargeable Battery Test Machine
1 . Sensitive galvanometer is used for students to check weak current or tiny voltage in DC circuit in the lab . It is a kind of high sensitive magnetoelectric instrument , which can measure the tiny current .
2 . dQ / dV differential capacity curve , the peak on the curve indicates that there is a voltage plateau on the charge and discharge curve , and different peaks represent different electrochemical reactions .
3 . The design of the uA current unit plays a vital role in the field of current accurate measurement , which is more convenient and more accurate .
4 . The lower range is able to output current low as 5uA , which can tell the subtle difference of various materials such electrodes , electrolyte , separator , etc .
5 . Generally , magnetoelectric galvanometer can achieve uA level , electromagnetic galvanometer can achieve mA level .
Input Power | 25W | |
Channel Features | Constant Current Source and Constant Voltage Source with Independent pairs of Closed - loop Structure | |
Channel Control Mode | Independent Control | |
Voltage | Per Channel Voltage Range | 0V - 5V |
Discharge Min Voltage | 0V | |
Accuracy | 0.05 percent of FS | |
Stability | 0.05 percent of FS | |
Current | Per Channel Current Range | Range1 , 5uA – 1mA Range2 , 1mA – 10mA Range3 , 10mA – 20mA |
Accuracy | 0.05 percent of FS | |
Stability | 0.05 percent of FS | |
Power | Per Channel Output Power | 0.25W |
Stop current | Range 1 , 2uA Range 2 , 0.02mA Range 3 , 0.04mA | |
Stability | 0.1percent of FS | |
Data Recording | Record Condition | Min Time - 100mS |
Min Voltage - 10mV | ||
Min Current - Range 1 , 2uA Range 2 , 0.02mA Range 3 , 0.04mA | ||
Record Frequency | 10Hz | |
Charge | Charge Mode | CCCharge , CVCharge , CCCVCharge , CPCharge |
End Conditon | Voltage , Current , Relative Time , Capacity , -deltaV | |
Discharge | Discharge Mode | CCD , CPD , CRD |
End Condition | Voltage , Current , Relative Time , Capacity |